| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) internal report # (b)(4).Reported defect not reproduced with returned meter and test strip.On qc investigation on 11/29/2018, returned meter and test strips produced e-7 results.Final root cause: rc-001 - miscellaneous user induced contamination on the strip connector.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with results obtained of 199 and 300mg/dl.The expected fasting blood glucose test result range is 120-140mg/dl.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored according to specification in the kitchen.The test strip lot manufacturer's expiration date is 04/24/2019 and open vial date is last month.The meter memory was reviewed for previous test result history.(b)(6).
|
| |
|
Search Alerts/Recalls
|