| Catalog Number |
50174180S |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available it will be provided on a supplemental report.Device dispositon is unknown.
|
| |
|
Event or Problem Description
|
|
As reported on medwatch (b)(4) : "during a surgical procedure removal external fixator of the right leg, removal of the frame the two anterior pins had fractured at the bone interface half of the pins was still remaining inside the femur, buried".Original intended procedure: removal of external fixator of the right leg.
|
| |
|
Event or Problem Description
|
|
As reported on medwatch 0501490000-2018-8007: "during a surgical procedure removal external fixator of the right leg, removal of the frame the two anterior pins had fractured at the bone interface half of the pins was still remaining inside the femur, buried".Original intended procedure: removal of external fixator of the right leg.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event that self-drilling half pin apex ha-coated ø 5mm, 180 x 40mm was alleged of 'breakage during surgery' could be confirmed, based on provided pictures.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.Based on provided pictures, we can only confirm the event: tip of the pin is broken.Further inspection is not possible without the returned device.
|
| |
|