Catalog Number 75446550
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Image review: pre-revision and post revision x-rays for thoraco lumbar stabilization construct provided.In two of the provided images, there appears to be broken screw at the inferior level of a construct around a burst fracture.Each of the screws is connected probably 2 degree to osteoporosis.The construct was revised to a compacting graft at the fracture level.Hardware failure was likely 2 degree to failure of fusion given osteoporosis.Product analysis: visual, microscopic and dimensional evaluation confirmed the bone screw disassembled from the head.The ring and crown were still assembled in the head.A portion of the ring has been sheared through the cross-sectional area, with the remaining portion of the ring still seated in the groove of the head.The retaining ring is fully seated but has been deformed and partially sheared off.The deformed and sheared areas are on either side of the retaining ring gap and would reduce the force required for head disassembly.The diameter of the bone screw head in the undamaged area was checked and meets print spec.The above observations are consistent with overload as the mechanism of failure.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent minimally invasive transforaminal lumbar interbody fusion.Post-op, the screw head and thread was found separated and broken.The product came in contact with patient.No fragments of the broken part remains in the patient.Patient complications were unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key8140015
Report Number1030489-2018-01608
Device Sequence Number18886258
Product Code KWP
UDI-Device Identifier00885074134130
UDI-Public00885074134130
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K042025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75446550
Device Lot NumberH5359336
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2018
Initial Report FDA Received Date12/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-