SMITH & NEPHEW, INC. LEGION PS OXIN FEM SZ8 LT; PROSTHESIS KNEE PATELLOFEMOROTIBIAL SEMI-CONSTRAINED CEMENTED POLY/METAL/POLY Back to Search Results
Catalog Number 71421218
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 09/12/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed in the patient due to pain and swelling.No delay reported.Implants replaced with competitor devices.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.The clinical/medical team concluded, based on the documentation provided, it appears the root cause of the patient¿s symptoms were multifactorial as the reported vmo atrophy, the right quad tendon tear, and loosening of components were all likely contributing factors leading to possible micro-motion, the subsequent instability, lateral patellar subluxation, swelling and pain.The patient impact was the reported persistent symptoms, bilateral tka revisions and post-surgical rehabilitation.No further medical assessment can be rendered at this time.A review of the production documentation for the corresponding products did not reveal any deviation from the standard manufacturing processes.A review of complaint history for the listed parts revealed no prior complaints for the listed batches/failure mode.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION PS OXIN FEM SZ8 LT
Common Device Name
PROSTHESIS KNEE PATELLOFEMOROTIBIAL SEMI-CONSTRAINED CEMENTED POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8140250
Report Number1020279-2018-02710
Device Sequence Number16616144
Product Code JWH
UDI-Device Identifier00885556031179
UDI-Public00885556031179
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K043440
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/10/2023
Device Catalogue Number71421218
Device Lot Number13CM07576
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/15/2018
Supplement Date Received by Manufacturer11/15/2018
Initial Report FDA Received Date12/06/2018
Supplement Report FDA Received Date06/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age58 YR
-
-