• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G53585
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
510 (k) number; k142688.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Lab evaluation the device related to this occurrence underwent a laboratory evaluation on 21 nov 2018.The needle was found to be broken proximal to the sheath extender.Document review prior to distribution, all echo-hd-19-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo-hd-19-c of lot number c1192809 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1192809.Ifu review the notes section of the instructions for use, ifu0077-4, which accompanies this device instructs the user to inspect the device prior to use : "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use (ifu0077-4 ).Root cause review a definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible cause could be attributed to the device being used in a flexed or twisted position during the procedure or tortuous anatomy as indicated in the additional information.Summary complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
Needle broke off."as per complaint form": after the punction, the needle was outside of the scope, the nurse will put out the needle to check the biopsy, the broken needle came out of the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8140805
Report Number3001845648-2018-00564
Device Sequence Number13690594
Product Code FCG
UDI-Device Identifier00827002535855
UDI-Public(01)00827002535855(17)190119(10)C1192809
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/19/2019
Device Model NumberG53585
Device Catalogue NumberECHO-HD-19-C
Device Lot NumberC1192809
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/06/2018
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2018
Initial Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-