• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Use of Device Problem (1670)
Patient Problem No Patient Involvement (2645)
Event Date 09/11/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 gas blender system.The incident occurred in (b)(6).The device was returned to livanova (b)(4) for further investigation.During the intense evaluation the reported issue could not be confirmed or reproduced.The device worked according the specification.The device was cleaned and calibrated and subsequent functional verification testing was completed without issues and the unit was returned to service.
 
Event Description
Livanova (b)(4) received a report that a s5 gas blender system was found to be defective during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key8141010
MDR Text Key129595782
Report Number9611109-2018-01381
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25-40-45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-