MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 5532-G-509
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 11/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
Due to a disengaged tibial insert, it was removed and replaced with a new one.
 
Event or Problem Description
Due to a disengaged tibial insert, it was removed and replaced with a new one.
 
Additional Manufacturer Narrative
An event regarding disassociation involving a triathlon insert was reported.The event was confirmed.Method & results: device evaluation and results: material analysis report indicates: material loss, material deformation and material wear were observed on distal condyle regions and the patella track; consistent with the insert disassociating from the baseplate.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: could not be performed as lot code information was unknown complaint history review: could not be performed as lot code information was unknown.Conclusions: the posterior end of the medial condyle had material loss consistent with the insert loaded from the posterior side.The material loss, material deformation and material wear observed on the distal surface of the insert, likely resulted from the insert disassociating from the baseplate.Yellow discoloration was also observed on the insert, consistent with absorption of synovial fluid.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.The event could be confirmed however, the root cause was not determined because insufficient medical information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.If additional information are received, this investigation will be reopened and re-evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8141694
Report Number0002249697-2018-03939
Device Sequence Number16197057
Product Code JWH
UDI-Device Identifier07613327050776
UDI-Public07613327050776
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number5532-G-509
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Initial Date Received by Manufacturer 11/11/2018
Supplement Date Received by Manufacturer02/12/2019
Initial Report FDA Received Date12/07/2018
Supplement Report FDA Received Date03/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age75 YR
Patient Weight115
-
-