• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT W/GUARDIVA & PROBE COVER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT W/GUARDIVA & PROBE COVER; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a kink in the stylet was confirmed, but the exact cause could not be determined from the four photographs that were provided for investigation.The photos showed one 3cg stylet that had been removed from the picc.The picc was not observed in the photos.The t-lock extension set remained attached to the stylet.A kink was observed in the stainless steel core wire between the t-lock extension set and the polyimide tubing.A kink was also observed in the polyimide tubing.The kink in the polyimide tubing appeared to be located proximal to the magnets.The stylet appeared complete and no part of the stylet appeared to be broken.Since the stylet was kinked, the complaint was confirmed; however the cause of the stylet could not be determined.Based on the condition of the stylet, it is possible that the stylet was damaged during use.The ifu for the 3cg stylet states, ¿ensure that the stylet tip does not extend beyond the trimmed end of the catheter.Extension of the stylet tip beyond the catheter end, combined with kinking and excessive forces, may result in vessel damage, stylet damage, difficult removal, stylet tip separation, potential embolism and risk of patient injury.¿ a lot history review (lhr) of recq0261 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (recq0261) have been reported from the same facility.
 
Event or Problem Description
It was reported that the stylet was kinked upon removal after successful picc placement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT W/GUARDIVA & PROBE COVER
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8142196
Report Number3006260740-2018-03502
Device Sequence Number1910163
Product Code LJS
UDI-Device Identifier00801741154904
UDI-Public(01)00801741154904
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2019
Device Model NumberN/A
Device Catalogue NumberS1295108FD2
Device Lot NumberRECQ0261
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2018
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/13/2018
Initial Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
Patient Weight65
-
-