| Model Number |
6227DEF04 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that a problem occurred during slitting of the catheter.During slitting of the deflectable catheter, the slitting blade encountered a wire.The physician stopped during slitting, made an adjustment and completed the slit without complications.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|