| Catalog Number |
PND6F0701058M |
| Medical Device Problem Codes |
Device Damaged Prior to Use (2284); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the neuron 070 was kinked approximately 57.5 cm, 102.0 cm and 103.5 cm from the hub.The device had bends approximately 29.0 cm and 86.0 cm from the hub.The device was ovalized from approximately 102.5 cm to 111.0 cm from the hub.The distal tip of the device was ovalized.Conclusion: evaluation of the returned neuron 070's confirmed damage along their length.If the devices are forcefully removed from their packaging at extreme angles or otherwise mishandled during preparation, damage such as this may occur.It is possible some of the returned damage on these devices was incidental to the reported complaint and may have occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-02382.
|
| |
|
Event or Problem Description
|
|
During preparation for a medical procedure, two neuron 6f 070 delivery catheters (neuron 070) were found to have bent tips.The damage to both neuron 070's was found prior to use, upon removal from the packaging and, therefore, were not used in the procedure.The procedure was completed using another catheter.
|
| |
|