PENUMBRA, INC. NEURON 6F 070 DELIVERY CATHETER; DQY Back to Search Results
Catalog Number PND6F0701058M
Medical Device Problem Codes Device Damaged Prior to Use (2284); Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the neuron 070 was kinked approximately 57.5 cm, 102.0 cm and 103.5 cm from the hub.The device had bends approximately 29.0 cm and 86.0 cm from the hub.The device was ovalized from approximately 102.5 cm to 111.0 cm from the hub.The distal tip of the device was ovalized.Conclusion: evaluation of the returned neuron 070's confirmed damage along their length.If the devices are forcefully removed from their packaging at extreme angles or otherwise mishandled during preparation, damage such as this may occur.It is possible some of the returned damage on these devices was incidental to the reported complaint and may have occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-02382.
 
Event or Problem Description
During preparation for a medical procedure, two neuron 6f 070 delivery catheters (neuron 070) were found to have bent tips.The damage to both neuron 070's was found prior to use, upon removal from the packaging and, therefore, were not used in the procedure.The procedure was completed using another catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON 6F 070 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8142729
Report Number3005168196-2018-02383
Device Sequence Number217152
Product Code DQY
UDI-Device Identifier00814548010052
UDI-Public00814548010052
Combination Product (Y/N)Y
Initial Reporter CountryCO
PMA/510(K) Number
K082290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/08/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2021
Device Catalogue NumberPND6F0701058M
Device Lot NumberF82298
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2018
Initial Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-