The reported device was not returned to olympus; therefore, the cause of the reported event could not be confirmed.However, if additional information becomes available or if the device is returned at a later date, this report will be supplemented accordingly.
|
|
Olympus was informed that at the end of a per oral pyloromyotomy (pop) procedure, the tip of the knife and a piece of the stem was observed missing from the device.An egd-esophagogastroduodenoscopy and fluoroscopy were performed to determine the location of distal end/knife and it added an additional 20-30 minutes under anesthesia for these extra procedures.The device fragments were retrieved from the patient's stomach using biopsy forceps.The intended procedure was completed and there was no patient injury reported.No devices or equipment were replaced.In addition, the device was visually inspected by the nurse and the physician prior to insertion into of device into an olympus 190 scope (model not specified); no anomalies were noted.Per the physician, the knife was noted to be slightly stiff to insert.However, there was no restriction of the scope and no difficulty withdrawing knife from scope.No sparking was observed.
|