MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6531
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/01/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.During preparation of an 12mm x 2.25mm nc quantum apex balloon catheter, the physician tried to insert the device into the guide catheter.While advancing the shaft got broke outside the patient body.No further complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a nc quantum apex balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the hypotube is separated 79.1cm from the hub.There are multiple kinks along the hypotube.Microscopic examination revealed no additional damages.There is blood present in the guidewire lumen and on the balloon folds.The balloon is still tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that shaft break occurred.During preparation of an 12mm x 2.25mm nc quantum apex balloon catheter, the physician tried to insert the device into the guide catheter.While advancing the shaft got broke outside the patient body.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8143445
Report Number2134265-2018-63570
Device Sequence Number1910159
Product Code LOX
UDI-Device Identifier08714729783237
UDI-Public08714729783237
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Model Number6531
Device Catalogue Number6531
Device Lot Number0018897826
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2019
Initial Date Received by Manufacturer 11/08/2018
Supplement Date Received by Manufacturer01/14/2019
Initial Report FDA Received Date12/07/2018
Supplement Report FDA Received Date02/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
-
-