| Model Number |
2477-0000 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 250ml baxter bag ndc 0338-0049-02, lot number v18g14a, exp jan20, 0.9% sodium chloride.The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a blood transfusion in the pediatric clinic, the set separated above the drip chamber which caused blood to spill.There's no report of patient harm.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a blood transfusion in the pediatric clinic, the set separated above the filtered drip chamber which caused blood to spill.There's no report of patient harm.The customer later reported this event "just happened last month.".
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report that the set separated above the drip chamber which caused a leak was confirmed.Visual inspection observed a separation between the top of the blood filter drip chamber (p/n 600093) where one of the two tubing lines (p/n 660134) enter.Closer inspection identified insufficient solvent at the engagement between the tubing and the top drip chamber ports.Dimensional analysis was within specification.The root cause of the customer's report was not determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: event.
|
| |
|
Event or Problem Description
|
|
The customer later reported this event ".Just happened last month.".
|
| |
|