OLYMPUS WINTER & IBE GMBH HF-CABLE, MONOPOLAR, 3.5 M; ELECTRICAL-ONLY MEDICAL DEVICE CONNECTION CABLE, REUSABLE Back to Search Results
Model Number A0336.1
Medical Device Problem Codes Melted (1385); Flare or Flash (2942)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported cable device was not returned to olympus for evaluation.The cause of the reported complaint cannot be confirmed.Despite several followup attempts, there is no further information regarding the procedure, the location of reported burning / melting on the device, the service age of the device, or the handpiece involved.As preventive measures, the cable instructions state, ¿do not use the hf cable after one year of use.¿ ¿visually inspect the cable and the plugs for irregularities on the surface.¿ and ¿do not use a cable with brittle or defective insulation.Replace the cable.¿ the instructions also warn, ¿use only compatible equipment¿ listed in the instructions, and ¿the maximum output voltage of the hf unit must not exceed the lowest maximum rated voltage of any of the hf components used during the procedure.¿ the user facility was also advised of the one year maximum usage recommendation at the time that the event was reported to olympus.
 
Event or Problem Description
Olympus was informed that at the end of an unspecified procedure, the cord had a flame and melted.There was no reported injury to the patient or to any other person.It was reported that the procedure was completed successfully since the event occurred at the end of the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-CABLE, MONOPOLAR, 3.5 M
Common Device Name
ELECTRICAL-ONLY MEDICAL DEVICE CONNECTION CABLE, REUSABLE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
MDR Report Key8143635
Report Number2951238-2018-00746
Device Sequence Number244107
Product Code FAS
UDI-Device Identifier04042761001472
UDI-Public04042761001472
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K951863
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberA0336.1
Device Catalogue NumberA0336.1
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2018
Initial Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-