Model Number 9770001
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Biomed reported system has poor ultrasound image quality.
 
Additional Manufacturer Narrative
Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of poor image quality was unconfirmed.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.The device was discarded due to being discontinued product.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Biomed reported system has poor ultrasound image quality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 6 PROBE
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8143795
Report Number3006260740-2018-03524
Device Sequence Number16672280
Product Code IYO
UDI-Device Identifier00801741125744
UDI-Public(01)00801741125744
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K071204
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 05/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9770001
Device Catalogue Number9770001
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2020
Event Location Hospital
Initial Date Received by Manufacturer 11/14/2018
Supplement Date Received by Manufacturer05/15/2020
Initial Report FDA Received Date12/07/2018
Supplement Report FDA Received Date05/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-