| Model Number |
TRUEMETRIX |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) internal report # 01604507 returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage(kitchen).Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(4) 2018, in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high control results.The customer is concerned with control tests results of 61mg/dl.The expected level 1 control solution range is 23-53mg/dl.Customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored according to specification in the kitchen.The test strip lot manufacturer's expiration date is 02/29/2020 and open vial date is 11/16/2018.The meter memory was not reviewed for previous test result history.Customer called for assistance running control test before using meter.Control result was out of range.
|
| |
|