LENSTEC BARBADOS INC. SOFTEC HD; INTRAOCULAR LENS Back to Search Results
Lot Number 181424
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
After thorough investigation by the quality systems department, the results of lenstec's investigation are as follows; lenstec can confirm that all procedures in the manufacturing and packaging of the lens was conducted correctly.The lens appeared to be handled whilst in a dehydrated state and this caused the break at the haptic.For this reason, lenstec recommends (following the dfu) that the lens should be implanted within two minutes of removal from the vial.Due to the hydrophilic nature of the lens, extended periods of time outside the saline will cause the lens to dehydrate and subsequently become damaged during the injection process.(b)(4).
 
Event or Problem Description
Lenstec, inc.Received an email notification stating "haptic broken during implantation, incision enlarged and lens removed.Another hd +17.0 was implanted to patient, closed with a stitch and no adverse event".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC HD
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church, BB170 92,
BB  BB17092,
MDR Report Key8144019
Report Number9613160-2018-00037
Device Sequence Number11240331
Product Code HQL
UDI-Device Identifier00844369027199
UDI-Public00844369027199
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/19/2023
Device Lot Number181424
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/09/2018
Initial Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-