| Model Number |
LF4318 |
| Medical Device Problem Codes |
Protective Measures Problem (3015); Intermittent Energy Output (4025)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The device knife blade was not exposed while the jaws were opened.The reported conditions were not confirmed.The jaws opened and closed properly when the handle was retracted and released.The knife blade advanced and retracted smoothly along the knife track.The knife cut of the device was tested on a silicone test strip with acceptable results.The knife blade was also inspected under microscope and no issue was found.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.The investigation found the device to function normally and within specifications.The instructions for use (ifu) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the total pelvic resection, at cutting process, the knife did not move and it was stuck without any movement.The knife, advanced and was exposed, after that it stopped.Surgery was completed by another device.No patient injury.
|
| |
|
Search Alerts/Recalls
|