MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE IMPACT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE IMPACT; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF4318
Medical Device Problem Codes Protective Measures Problem (3015); Intermittent Energy Output (4025)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned product met specification as received.Visual inspection found no defects.The device knife blade was not exposed while the jaws were opened.The reported conditions were not confirmed.The jaws opened and closed properly when the handle was retracted and released.The knife blade advanced and retracted smoothly along the knife track.The knife cut of the device was tested on a silicone test strip with acceptable results.The knife blade was also inspected under microscope and no issue was found.The device was activated ten times on a saline soaked towel with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.The device was activated with the rotation knob in various positions to detect any problematic activation issues.The investigation found the device to function normally and within specifications.The instructions for use (ifu) states, keep the instrument jaws clean.Build up of eschar may reduce sealing and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during the total pelvic resection, at cutting process, the knife did not move and it was stuck without any movement.The knife, advanced and was exposed, after that it stopped.Surgery was completed by another device.No patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE IMPACT
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key8144456
Report Number1717344-2018-01605
Device Sequence Number10828720
Product Code GEI
UDI-Device Identifier10884521165311
UDI-Public10884521165311
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K123444
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2023
Device Model NumberLF4318
Device Catalogue NumberLF4318
Device Lot Number80610222X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2018
Initial Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
Patient Weight72
-
-