• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MSD DEGGENDORF MFG PEEK PREVAIL CERVICAL INTERBODY DEVICE; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, CERVICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MSD DEGGENDORF MFG PEEK PREVAIL CERVICAL INTERBODY DEVICE; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, CERVICAL Back to Search Results
Catalog Number 4210564
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was pre-operatively diagnosed with cervical degenerative disease/ spinal stenosis and underwent anterior cervical decompression.During the surgery, a fracture occurred during the partial implantation of peek.The director used a pair of pliers to clamp it out.The procedure was completed successfully with a new product.No patient complications occurred as a result of the event.
 
Additional Manufacturer Narrative
Product analysis result: visual optical visual and optical inspection confirmed the peek implant is broken.The top portion of the implant has been broken off.Optical inspection confirmed a flat fracture surface with no indication of pre existing failure that could contribute to the implant breaking.Optical inspection confirmed some deformation on the ribs of the implant.The threads at the inserting point of the screw appear to be damaged.Unable to determine root cause of failure.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEEK PREVAIL CERVICAL INTERBODY DEVICE
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, CERVICAL
Manufacturer (Section D)
MSD DEGGENDORF MFG
wertstrasse 17
deggendorf 94469
MDR Report Key8144587
Report Number1030489-2018-01616
Device Sequence Number10811783
Product Code ODP
UDI-Device Identifier00613994502087
UDI-Public00613994502087
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K073285
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2024
Device Catalogue Number4210564
Device Lot Number21CW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2019
Initial Date Received by Manufacturer 11/09/2018
Supplement Date Received by Manufacturer04/19/2019
Initial Report FDA Received Date12/07/2018
Supplement Report FDA Received Date04/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-