• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. NEURON 6F 070 DELIVERY CATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. NEURON 6F 070 DELIVERY CATHETER; DQY Back to Search Results
Catalog Number PND6F0701058M
Medical Device Problem Codes Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the returned neuron 070 had kinks approximately 4.0 cm, 33.0 cm and 97.0 cm from the hub.The device had ovalizations along the distal shaft approximately from 98.0 cm to 111.0 cm from the hub.Conclusions: evaluation of the returned neuron 070 revealed an ovalization on the distal tip.If the device is forcefully gripped or otherwise mishandled, the device may become ovalized.This ovalization likely contributed to the inability to advance a guidewire through the subject device.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a medical procedure using a neuron 6f 070 delivery catheter (neuron 070).During the procedure, the physician experienced resistance while advancing the neuron 070 through the sheath.The physician was then unable to make a pass with a guidewire through the neuron 070, therefore; the neuron 070 was removed and found to be ovalized.The procedure was completed using another catheter with the same sheath.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON 6F 070 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8144687
Report Number3005168196-2018-02381
Device Sequence Number13859111
Product Code DQY
UDI-Device Identifier00814548010052
UDI-Public00814548010052
Combination Product (Y/N)Y
Initial Reporter CountryCO
PMA/510(K) Number
K082290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/08/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/12/2021
Device Catalogue NumberPND6F0701058M
Device Lot NumberF82298
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2018
Initial Report FDA Received Date12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Date Report Sent to FDA01/01/2005
-
-