| Model Number |
9733449 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information was unavailable from the site.The device was returned to the manufacturer for evaluation.Functional testing; visual/physical examination was performed and the issue was confirmed.The returned straight suction was not able to verify when attached to a known good tracker returning a divot error of 4.4 mm to 4.7 mm.The cause of the issue was due to a damaged tool.Device manufacturing date is unknown.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess).It was reported that during the case the straight suction was found to be damaged.A new suction was used to completed the procedure.There was no surgical delay due to this issue, and there was no impact on patient outcome.
|
| |
|