• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523RNAS
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 09/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they experienced low blood glucose.The customer¿s blood glucose level was in the range of high 50 mg/dl, 60 mg/dl and in the range of 60 mg/dl to 70 mg/dl.The current blood glucose level is 111 mg/dl.The customer treated for low blood glucose value with food.Based on customer¿s report customer allege the insulin pump was over delivering reason that they received a reminder of a missed bolus after delivering a bolus and then went low.The customer stated that the drive support cap was normal.The customer was neither in the emergency room, nor admitted into hospital as a result of low blood glucose.The customer was wearing the insulin pump when low blood glucose event was reported.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump passed the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test.No device deficiency anomaly noted during test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8145895
Report Number2032227-2018-72879
Device Sequence Number11189040
Product Code OYC
UDI-Device Identifier00613994748706
UDI-Public(01)00613994748706
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523RNAS
Device Catalogue NumberMMT-523RNAS
Device Lot NumberA000167945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2019
Initial Date Received by Manufacturer 11/29/2018
Supplement Date Received by Manufacturer05/08/2019
Initial Report FDA Received Date12/10/2018
Supplement Report FDA Received Date05/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRN-UNK-RSVR UNOMED SET
Outcome Attributed to Adverse Event Other;
Patient Age78 YR
Patient Weight51
-
-