• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY UNI TIBINRT S5-6RM/LL8MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY UNI TIBINRT S5-6RM/LL8MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71422265
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/31/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed one year after original surgery and insert showed delamination.
 
Additional Manufacturer Narrative
The associated journey uni tibial insert was returned and evaluated.A lab analysis conducted during this investigation indicated that the articulating surface of the insert showed wear and deformation.There were no observations of material or manufacturing deviations in the course of this investigation.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A clinical evaluation noted that based on the explant analysis and without pre and post-op imaging for comparison the root cause of the insert scratches, reported early delamination and intraoperative findings cannot be concluded.However, the ¿slight projection of the tibial plateus to the medial (approx.1 mm)¿ from the primary uni-implantation, the small osteophyte (which was removed during the revision), and/or possible 3rd body debris could have been contributing factors to the reported event.The patient impact beyond the wear, revision, and expected transient post-op convalescence period cannot be determined.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY UNI TIBINRT S5-6RM/LL8MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8146232
Report Number1020279-2018-02722
Device Sequence Number1279686
Product Code HSX
UDI-Device Identifier03596010582249
UDI-Public03596010582249
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K061011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Expiration Date09/20/2026
Device Catalogue Number71422265
Device Lot Number16JM12003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2019
Initial Date Received by Manufacturer 11/13/2018
Supplement Date Received by Manufacturer11/13/2018
Initial Report FDA Received Date12/10/2018
Supplement Report FDA Received Date04/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-