| Model Number |
2781 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
11/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Initial reporter phone: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the guidewire broke and a piece remained inside the patient.A 300cm pt2 guidewire was selected for use for a procedure in the lower leg.The patient had grade 4 calcification on the lower leg.The wire broke as the doctor passed the stenosis.The piece of wire remained in the patient and could not be recovered.According to the physician, the strong calcification was to blame for breaking the wire.
|
| |
|
Event or Problem Description
|
|
It was reported that the guidewire broke and a piece remained inside the patient.A 300cm pt2 guidewire was selected for use for a procedure in the lower leg.The patient had grade 4 calcification on the lower leg.The wire broke as the doctor passed the stenosis.The piece of wire remained in the patient and could not be recovered.According to the physician, the strong calcification was to blame for breaking the wire.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr: analysis of returned product revealed that the device is broken at the distal section.Only 298.5 cm was returned.Also the distal end is kinked at the broken section.
|
| |
|