STRYKER ORTHOPAEDICS-MAHWAH MOD PLASMA HA STM 11MM X 127MMRESTORATION MODULAR HIP SYSTEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6276-5-011
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 11/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.
 
Event or Problem Description
It was reported that patient's left hip was revised due to dislocation.The head dislocated from the liner as a result of the proximal body disassociating from the plasma stem and pulling the head out of the liner.Patient had been performing a non-compliant activity (bending over) when dislocation/ disassociation occurred.Surgeon reported he was unsure if the disassociation occurred because of surgeon error or device failure of the locking mechanism.Rep noted that during implantation, surgeon did not use torque wrench, but applied force by free-hand tightening the locking mechanism.A biolox ceramic head and restoration proximal body were revised to the same catalog numbers of different lot codes as the surgeon decided not to re-implant the revised devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOD PLASMA HA STM 11MM X 127MMRESTORATION MODULAR HIP SYSTEM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8146433
Report Number0002249697-2018-03948
Device Sequence Number1279689
Product Code LPH
UDI-Device Identifier04546540641809
UDI-Public04546540641809
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Catalogue Number6276-5-011
Device Lot Number61989704
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2018
Initial Report FDA Received Date12/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;