MASIMO - 40 PARKER O3 REGIONAL OXIMETER MODULE Back to Search Results
Model Number 24667
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported the o3 oximeter is giving low rso2 values.No patient impact or consequences were reported.
 
Event or Problem Description
The customer reported the o3 oximeter is giving low rso2 values.No patient impact or consequences were reported.
 
Additional Manufacturer Narrative
Additional manufacuring narrative: other, other text: the returned device was evaluated.Visual inspection upon receiving revealed no external damage or defects.Sensor was able to obtain measurements on each channel at 75% rso2 through 10 minutes of continuous testing; measurements were acceptable when compared to measurements obtained by a known good o3 sensor.Internal inspection revealed contaminant on u2 chip on analog board.The reported issue was not duplicated as the sensor provided accurate readings.However, the contaminant on analog board u2 chip could potentially cause the lower values seen by customer.A service history record review reveals that this unit was in the field for over one (1) year with no previous reported issues prior to this reported event.Model # was corrected from 9637 to 24667.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O3 REGIONAL OXIMETER MODULE
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8146667
Report Number2031172-2018-00566
Device Sequence Number11423264
Product Code MUD
UDI-Device Identifier00843997010771
UDI-Public00843997010771
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K160526
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24667
Device Catalogue Number9637
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/19/2018
Supplement Date Received by Manufacturer01/27/2019
Initial Report FDA Received Date12/10/2018
Supplement Report FDA Received Date02/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-