| Model Number |
CTS22 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event device has been returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: robot assisted prostatectomy.Event description: the dr.Wants to see the placing of the second trocar through the camera inserted in the first trocar, this leads to a very limited space and the trocars collide with each other.This occurs almost every time but this is the first time a trocar has broken from it.The piece that broke off remained on the trocar and came out along with it when they changed if for a new one.Additional information received from the sales rep by email on 22 nov 18: i wouldn't call it a clean break.The tip was bent inwards and the piece that broke off didn't break off immediately.The break was at the very tip of the cannula.Patient status: nothing "happend" to the patient as the piece of plastic remained attached to the trocar and it was changed immediately.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.The complainant's experience of a fragmented cannula tip was not confirmed, as engineering observed that the cannula tip was bent inward with no pieces missing.Based on the condition of the unit that was returned, the exact root cause of the bent cannula tip could not be determined.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.This event is not reportable as it is unlikely to cause or contribute to death or serious deterioration in the state of health.
|
| |
|
Event or Problem Description
|
|
Procedure performed: robot assisted prostatectomy.Event description: the dr.Wants to see the placing of the second trocar through the camera inserted in the first trocar, this leads to a very limited space and the trocars collide with each other.This occurs almost every time but this is the first time a trocar has broken from it.The piece that broke off remained on the trocar and came out along with it when they changed if for a new one.Additional information received from the sales rep by email on 22 nov 18: i wouldn't call it a clean break.The tip was bent inwards and the piece that broke off didn't break off immediately.The break was at the very tip of the cannula.Additional information received from the sales rep by email on 03 jan 2019: the second trocar that was used at the time of this event was damaged and returned for evaluation, not the first trocar used for visualization with the scope.They believe that the trocar where the tip broke off hadn't been connected to the robot yet.The breaking occurred during the insertion of the trocars.The optics is they one the used for robot surgery, 12mm 0 degrees.Patient status: nothing happend to the patient as the piece of plastic remained attached to the trocar and it was changed immediately.
|
| |
|