Additional Manufacturer Narrative
Based on the descriptions of the event by user facility, the cause of no flow is probably due to poor connection with connector used on patient side.A tight connection is needed to open the fluid channel in the valve of some patient connectors.The pump flows normally when disconnected.
 
Event or Problem Description
Smartez 100 ml volume, 200 ml / hr, filled with ceftriaxone 2 grams/0.9% sodium chloride 100 ml, would not infuse through iv line, but would drip when disconnected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTEZ PUMP
Common Device Name
ELASTOMERIC INFUSION PUMP
Manufacturer (Section D)
EPIC INTERNATIONAL (THAILAND) CO., LTD.
hemaraj eastern seaboard indus
500/73 moo 2, tasit
pluakdaeng, rayong 21140
TH  21140
MDR Report Key8147452
Report Number3012429465-2018-00003
Device Sequence Number16536471
Product Code MEB
UDI-Device Identifier28859299101013
UDI-Public(01)28859299101013
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K151650
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number481012
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2018
Initial Report FDA Received Date12/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-