• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® BUR; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® BUR; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1883672HS
Device Problems Use of Device Problem (1670); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Analysis found that the inner shaft was bent until broken, the spiral wrap was stretched, the irrigation barb was deformed, and there were tool marks along the length of the bur.The outside diameter of the inner hub showed deformation.The inner hub outside diameter requirement is 0.330¿ / +-0.002¿ and measured 0.329¿ in the undamaged area and 0.362¿ in the deformed area which is out of specification.The hub bushings were gouged as well.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A health care provider (hcp) reported that the device was not working properly.The bur was stuck in the shaver prior to the procedure.There was no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
uriza shums
6743 southpoint drive north
jacksonville, FL 32216
9043328405
MDR Report Key8147515
MDR Text Key130048938
Report Number1045254-2018-00694
Device Sequence Number1
Product Code EQJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1883672HS
Device Catalogue Number1883672HS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2018
Initial Date FDA Received12/10/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-