| Model Number |
1863 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned sensor was evaluated.During evaluation the sensor passed all visual, continuity, and functional testing.The sensor fit firmly on fingers.No physical damage was observed.No short or open connection was detected, and no intermittent short or open connection was detected when cable was manipulated.The sensor passed spo2 simulation tests and was able to obtain spo2 (96%) and pulse rate (75 bpm) values.
|
| |
|
Event or Problem Description
|
|
The customer reported the sensor is too loose and the oxygen status is not correct.No patient impact or consequences were reported.
|
| |
|