| Model Number |
9731780 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No parts have been received by the manufacturer for evaluation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system.It was reported outside of a procedure that there were some instruments that were damaged and need to be replaced.No patient was present at the time of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument udi, 510k # and manufacturer date were updated to the returned instrument.The open spine clamp was returned to the manufacturer for evaluation.After functional testing and visual/physical examination the reported issue was confirmed.The adjustment screw threads were damaged and there were tool marks around the head of the screw of the returned clamp.Physical damage was observed.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system.It was reported outside of a procedure that there were some instruments that were damaged and need to be replaced.No patient was present at the time of the event.
|
| |
|