| Catalog Number |
3542610 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); No Code Available (3191)
|
| Date of Event |
11/09/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A device history record review is in progress.Once completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) patient underwent a primary breast augmentation procedure with a mentor siltex round moderate profile 125cc saline breast prosthesis that deflated after implantation.Deflation was on the patient¿s right side.In addition, the patient developed capsular contracture (baker grade i) in her right breast.As a result, the patient underwent explantation and replacement with a mentor memorygel breast implant 175cc gel breast prosthesis on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
On 02/15/2019, mentor completed the investigation on the suspect medical device.Upon receipt by mentor, the device contained no fluid.Yellow material was observed on the shell surface.During initial evaluation a rent was observed on the posterior aspect, measuring approximately 1.6 cm, however microscopic examination of the edges of the rent gave no indications of instrument damage.Deflation complaint was confirmed.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.Mentor product analysis lab concluded that the rent occurred sometime subsequent to the removal of the device from its protective packaging.The complaint was confirmed since a rupture was found on the device.Nonetheless, a microscopic examination of the edges of the rent was performed and it did not provide conclusive evidence of what could be the root cause.At this time is not possible to determine the origin of the yellow material observed.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet; however, deflation trends for mentor saline-filled devices will continue to be monitored by quality assurance.A complaint of capsular contracture in the breast was reported.With the information provided, the product evaluation (pe) team concluded the reported complaint of capsular contracture in the patient was the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture and deflation are known complications associated with these devices and are referenced in our current product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 02/04/2019, the mentor product analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2019, the device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|