| Catalog Number |
3549314 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Initial reporter phone number: (b)(6).Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) female patient underwent a breast reconstruction with a mentor cpx 4 breast tissue expander with suture tabs 550cc and experienced deflation on the left side.Deflation was confirmed by physician during exam.As a result, the patient underwent removal and replacement with an unspecified tissue expander on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: upon receipt by mentor, the device returned without fluid.No foreign material was observed within the device.White and brown material was observed on the shell surface.Upon initial inspection it was noticed a rent measuring approximately < 0.1 cm on the anterior aspect.Microscopic examination was performed.No evidence of instrument damage was observed.No other anomalies were discovered.The complaint was confirmed since a rupture was found on the device.However, at this point it is not possible to determine the root cause of such damage or the etiology of the white and brown material found on the device.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 1/13/2019, the mentor failure analysis lab received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|