Catalog Number 5901-1103
Health Effect - Clinical Code No Information (3190)
Date of Event 11/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
Upon incoming inspection, item was visibly worn.No further information is available.Broken tip at the threads.
 
Additional Manufacturer Narrative
The reported event that reunion tsa - version ro d was alleged of 'instruments - broken, deformed, worn or scratched' could be confirmed.Based on investigation, the root cause was attributed to be wear related.The failure was caused by the extensive use of the device.The device inspection revealed the following: the tip of the device seems broken from the pictures sent to complaint handling.The pictures provided show that the device has multiple scratches and the material color is faded, which proved that the device was extensively used.As a reminder, the cleaning and sterilization guide states: "note: stryker trauma & extremities typically does not specify the maximum number of uses appropriate for re-usable medical devices.The useful life of these devices depends on many factors including the method and duration of each use, and the handling between uses.For devices that may be impacted check that the device is not damaged to the extent that it malfunctions or that burrs have been produced that could damage tissues or surgical gloves.Careful inspection and functional test of the device before use is the best method of determining the end of serviceable life for the medical device.See appendix 2 for further details." a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
 
Event or Problem Description
Upon incoming inspection, item was visibly worn.No further information is available.Broken tip at the threads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REUNION TSA - VERSION RO D
Common Device Name
SHOULDER JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS.
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key8148507
Report Number0008031020-2018-01040
Device Sequence Number11185958
Product Code KWS
UDI-Device Identifier07613327099409
UDI-Public07613327099409
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K130895
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5901-1103
Device Lot Number6800701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2018
Initial Date Received by Manufacturer 11/12/2018
Supplement Date Received by Manufacturer01/17/2019
Initial Report FDA Received Date12/10/2018
Supplement Report FDA Received Date02/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-