• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US ES2 CANNULATED MODULAR TAP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US ES2 CANNULATED MODULAR TAP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 48280145
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Injury (2348); Foreign Body In Patient (2687)
Date of Event 11/20/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that jamshidi broke intra-op and 40mm of it stuck in the patient pedicle.The pedicle hole was enlarged to remove stuck jamshidi.As per email communication, tap was used and tip of it broken intra-op.Majority of the broken fragments were retrieved.A surgical delay of 2 hours was reported.
 
Additional Manufacturer Narrative
Method: nc/capa history review, labelling review, risk assessment result: the reported event was confirmed via email correspondence.Device was not returned and lot # was not provided, therefore device evaluation and product history review could not be performed.Conclusion: without device evaluation and product information, root cause could not be determined.Device not returned.
 
Event or Problem Description
It was reported that jamshidi broke intra-op and 40mm of it stuck in the patient pedicle.The pedicle hole was enlarged to remove stuck jamshidi.As per email communication, tap was used and tip of it broken intra-op.Majority of the broken fragments were retrieved.A surgical delay of 2 hours was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ES2 CANNULATED MODULAR TAP
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key8149016
Report Number0009617544-2018-00295
Device Sequence Number11073820
Product Code NKB
UDI-Device Identifier04546540710413
UDI-Public04546540710413
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K140400
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number48280145
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/20/2018
Supplement Date Received by Manufacturer01/31/2019
Initial Report FDA Received Date12/10/2018
Supplement Report FDA Received Date02/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-