| Model Number |
MICL13.2 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Optical Problem (3001)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); Halo (2227); No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
10/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -10.5 diopter, into the patient's eye.The surgeon reports the lens "is too big" and is planning to exchange it for a shorter lens.However, unable to obtain additional information despite repeated requests.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the lens was implanted in the left (os) eye.On (b)(6) 2018 the lens was exchanged with a shorter lens.The lens was returned in liquid and taped to an email.Visual inspection found no visible damage to the lens.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: lens work order search was performed.One similar complaint type event within associated lots found.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Event-updated event statement.Please disregard all previous event statements.Patient codes added.Device code added.Claim #(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that on (b)(6)2018 the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -10.5 diopter, into the patient's right (od) eye.On (b)(6)2018 the lens was exchanged with a shorter lens due to excessive vault and glare/halos.The problem was resolved.The cause of the event is reported as user error, stating "the implant placed was too big.".
|
| |
|