| Model Number |
7171/65 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
Heart Failure (2206)
|
| Date of Event |
11/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was admitted to the hospital for heart failure.Upon interrogation of the device, reproducible noise was noted on the right ventricular lead.The device had a history of noise in 2017 and was reprogrammed.The patient was stable.No additional information was reported.
|
| |
|
Event or Problem Description
|
|
New information noted that the noise occurred after atrial pacing.The device was programmed to vvi.This resolved the noise.
|
| |
|