| Catalog Number |
7714015 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Pain (1994); Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.X-ray result: intra-op films from revision surgery c4-6 acdf show fracture of plate and screws at c4-5.It is unclear if an interbody graft is present at c4-5 or what the fusion status at c5-6 is.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Type of procedure or technique used: anterior cervical discectomy and fusion (acdf) levels implanted: c4-6 it was reported that post-op, the implanted screw broke which resulted in non union.Fragments including distal end of screw that broke off remained inside the patient.Patient suffered neck/arm pain and the patient underwent revision surgery for the removal of implant.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis result: visual microscopic visual and microscopic inspection confirmed the screw has broken approximately 3 threads down from the screw head.Microscopic inspection revealed a fairly flat fracture surface with a circular material flow and the head of the screw has been stripped.This type of damage is consistent with torsional overload.The broken portion of the screw was not returned for analysis.If information is provided in the future, a supplemental report will be issued.
|
| |
|