| Catalog Number |
3003053 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Pain (1994); Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.X-ray result: intra-op films from revision surgery c4-6 acdf show fracture of plate and screws at c4-5.It is unclear if an interbody graft is present at c4-5 or what the fusion status at c5-6 is.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Type of procedure or technique used: anterior cervical discectomy and fusion (acdf) levels implanted: c4-6 it was reported that post-op, the implanted plate broke which resulted in non union.Fragments including half of the plate, remained inside the patient.Patient suffered neck/arm pain and the patient underwent revision surgery for the removal of implant.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis result: visual optical microscopic visual and optical inspection confirmed the plate is broke in multiple pieces.According to the event description the remaining piece is still implanted.Microscopic inspection confirmed a fairly flat fracture surface with fracture lip toward the edges of the break.There no evidence of damage that would suggest a defect in manufacturing of the implant.This type of damage is consistent with bend stress overload.If information is provided in the future, a supplemental report will be issued.
|
| |
|