• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAP
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/29/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform the displacement test due to completely broken reservoir tube lip.Unit had missing reservoir tube o-ring.
 
Event or Problem Description
The customer reported via phone call that insulin pump had damage in reservoir compartment.The customer's blood glucose was unknown at the time of incident.The customer reports that clear lip pops off concerned about losing security.Reservoir was not able to locked in place.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8149837
Report Number2032227-2018-73279
Device Sequence Number18899013
Product Code OYC
UDI-Device Identifier00643169446410
UDI-Public(01)00643169446410
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAP
Device Catalogue NumberMMT-523NAP
Device Lot NumberB3523NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2018
Initial Report FDA Received Date12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age25 YR
-
-