MAUDE Adverse Event Report: ZIMMER GMBH BIOLOX DELTA CERAMIC FEMORAL HEAD M 40/0, TAPER 12/14

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH BIOLOX DELTA CERAMIC FEMORAL HEAD M 40/0, TAPER 12/14 Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/24/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medical product : avenir, stem, standard, cemented, 5, taper 12/14 catalog #: 0106010205 ; lot #: 2818098, allofit it alloclassic shell for acetabulum, uncemented, 56/kk catalog #: 00875505600; lot #: 2817991.The manufacturer did receive x-rays, or other source documents for review.The manufacturer did not receive the device for investigation.Where lot numbers were received for the explanted devices, the device history records were reviewed and found to be conforming.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event or Problem Description
Patient was implanted on the right side and underwent medical intervention due to dislocation and pain.
 
Event or Problem Description
Please refer to report 0009613350-2018-01198.
 
Additional Manufacturer Narrative
This follow-up report is being filled to relay additional information, which was unknown at the time of the previous medwatch.Additional and corrected information are filled in the following fields: additional: weight - common device name - uf follow up, what type? correction: date of report - other relevant history - pma/510(k) - adverse event - additional mfg narrative.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available, an updated report will be submitted.Zimmer biomet¿s reference number of this file is cmp-(b)(4).
 
Additional Manufacturer Narrative
Investigation results were made available.Additional: correction: d11 - medical product: item: unknown pe insert item#: unknown lot#: unknown item: avenir, stem, standard, cemented, 5, taper 12/14 item#: 0106010205 lot#: 2818098 item: allofit it alloclassic, shell for acetabulum, uncem.56/kk item#: 00875505600 lot#:2817991.As the case at hand is a legal claim it is not suspected that the devices or additional information are being submitted for review.Zimmer gmbh winterthur legal department have already passed all information that was received from the lawyer, to our complaint handling department.By experience zimmer gmbh never gets more information except for the one that has been already covered in the final report.Patients¿ advocates only provide to zimmer gmbh as much information as they are willing to share to protect the rights of their clients.All information which has been provided for this particular case is already covered in the final report.Nevertheless, should additional information become available to us, a follow up report will be submitted.A technical investigation was not possible to be performed, as the device(s) were not at hand for investigation.However, based on the available information the investigation is conducted with outcome as follows.Dhr review: ref:00-8775-040-02 ; lot:2806914 - yield: (b)(4) - delivered:(b)(4) the device manufacturing quality records indicate that the released components met all requirements to perform as intended ref:0106010205 ; lot:2818098 - yield:(b)(4) - delivered:(b)(4) - scrapped:2 - reason for scrapping: unknown reasons the device manufacturing quality records indicate that the released components met all requirements to perform as intended ref: 00-8755056-00; lot:2817991 - yield:(b)(4) - delivered:(b)(4) - scrapped:5 - reason for scrapping:bore diameter is outside tolerance and outer diameter is outer tolerance the device manufacturing quality records indicate that the released components met all requirements to perform as intended.Trend analysis: a trigger considering the following event is identified: dislocation event description: it was reported that despite the revision surgery done on (b)(6) 2016, the patient felt pains.He is admitted to emergency which suspect dislocation of the hip prosthesis.Review of received data some legal documentation and patient visit reports were received (in french).The legal documentation mentions that the patient had 4 times dislocation within 5 months.All these dislocations were occurred after the first revision surgery done on (b)(6) 2016.No other relevant information in the legal documentation.The patient visit reports show that on (b)(6) 2016 during a medical consultation presence of foreign body was noticed: two metal splinters seen on x-rays, which were not seen post-operatively.On (b)(6) 2016 those foreign bodies seemed to belong to the insert which had migrated; a loss of thickness of the implant (compared to x-rays) is also noticed / mentioned.No anomalies in the interface between bone and cup were seen.Soft tissue with these fragments was taken to analyses.It is also noted that the patient had always pain after the first revision surgery and therefore got some painkiller.This patient had the initial surgery on (b)(6) 2015.After that time the patient had 2 revision surgeries and 4 times dislocation.See below the summary: (b)(6) 2016 1st revision surgery due to ceramic insert fracture (b)(6) 2016: 1st dislocation is suspected which is captured under (b)(4) (this complaint); 2nd dislocation occurred on (b)(6) 2016 and captured under (b)(4); 3rd dislocation occurred (b)(6) 2016 and captured under (b)(4); 4th dislocation occurred (b)(6) 2016 and captured under (b)(4) (b)(6) 2016 2nd revision surgery at the (b)(6) and captured under (b)(4) devices analysis: no product was returned to zimmer biomet for in-depth analysis.Review of product documentation: this device is intended for treatment./ all involved devices are intended for treatment.The compatibility check was performed and showed that the product combination was approved by zimmer biomet.Conclusion summary it was reported that despite the revision surgery done on (b)(6) 2016, the patient felt pains.He is admitted to emergency which suspect dislocation of the hip prosthesis.Some legal documentation and patient related reports were received (in french).These documents show that the patient had 2 revision surgeries and 4 dislocations within 5 months.In total 6 complaints were created for this patient.This complaint is for the first dislocation.Neither x-rays nor devices or photos of the explanted implants were received; therefore the condition of the components is unknown.The investigation results did not identify a non-conformance or a complaint out of box (coob).However, based on the given information and the results of the investigation, we were not able to identify a specific root cause for this issue.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet's reference number of this file is (b)(4).Note: the following report is associated with this event (same patient): (b)(4).
 
Event or Problem Description
Investigation results are now available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA CERAMIC FEMORAL HEAD M 40/0, TAPER 12/14
Common Device Name
BIOLOX DELTA CERAMIC FEMORAL HEAD
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
MDR Report Key8149898
Report Number0009613350-2018-01198
Device Sequence Number18893112
Product Code LZO
UDI-Device Identifier00889024430402
UDI-Public00889024430402
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K071535
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source consumer,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2025
Device Model NumberN/A
Device Catalogue Number00-8775-040-02
Device Lot Number2806914
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/13/2018
Supplement Date Received by Manufacturer03/29/2019
07/02/2019
Initial Report FDA Received Date12/11/2018
Supplement Report FDA Received Date04/16/2019
07/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE; SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight110
-
-