• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC KYPHON HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC KYPHON HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Catalog Number C01B
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the r eported event.X-ray result: post op ct shows extravasation of cement from vertebral body into the peri spinal plexus and pvc.No ct of chest is provided to r/o pulmonary or arterial embolism.This is a known complication of kyphoplasty.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent balloon kyphoplasty surgery at l2 due to compression fracture of l2 vertebral body.Post-op, the cement leaked outside of the vertebral body.The patient had a slight arrythmia that had resolved.Patient still had slight rib pain that the surgeon believes could be from positioning on to the operation table.Doctor was concerned about preventing cement from breaking off and causing a pulmonary embolism or travelling into the heart.From ct scan it was found that the cement had leaked outside of the vertebral body and travelled about 4cm inside the vena cava.Patient symptoms resolved and the patient is having no other symptoms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON HV-R BONE CEMENT
Common Device Name
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key8149899
Report Number1030489-2018-01627
Device Sequence Number13845313
Product Code NDN
UDI-Device Identifier00643169097834
UDI-Public00643169097834
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K180700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2021
Device Catalogue NumberC01B
Device Lot Number0009361275
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2018
Initial Report FDA Received Date12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
-
-