| Model Number |
18176550 |
| Medical Device Problem Code |
Off-Label Use (1494)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The pedicle screw involved in this fixation failure, as well as the set screw loosened, have been examined to check if they have been degraded during the construct tightening.The dimensional analysis conducted demonstrates that both pedicle screw and set screw are in conformity with the specifications of the technical drawings.The visual examination does not indicate any abnormal surface issue or material issue.
|
| |
|
Event or Problem Description
|
|
Two weeks post-op on a t10-l3 fusion construct for vertebral fracture treatment, the pedicle screw on contra-lateral side was pulled out of pedicle.As a consequence, the set screw loosened.The patient needed a revision surgery to resecure the construct.The construct was lengthened to l4 and pedicle screws which pulled out were replaced.The revision surgery was completed without further complications.
|
| |
|