| Catalog Number |
AB35W07200135 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician attempted to use an evercross balloon to treat a severely calcified lesion with cto (chronic total occlusion-100%) in the mid- superficial femoral artery (sfa).The vessel presented moderate tortuosity.Ifu was followed and the device was prepped with no issues noted.It was reported that a pin hole balloon burst during inflation to 8 atm.All fragments were retrieved from the patient.The procedure was completed with another balloon.No injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: contrast and heparinised saline were used to inflate the balloon.No issues were noted when withdrawing the balloon from the patient.The procedure was completed by carrying out a planned stenting and then ballooning.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|