| Catalog Number |
601220 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
Customer reported that while troubleshooting failed load cell calibrations it was discovered that a portion of the power cable appeared to be blackened and possibly burnt.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The device involved was returned for evaluation.The reported complaint was not confirmed.The load cell cable has normal cosmetic crimp marking.If additional pertinent information becomes available a follow-up report will be filed.
|
| |
|
Event or Problem Description
|
|
As per user facility: customer reported that while troubleshooting failed load cell calibrations it was discovered that a portion of the power cable appeared to be blackened and possibly burnt.No patient involvement.
|
| |
|
Search Alerts/Recalls
|