| Model Number |
V5922 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been recorded in leventon(b)(4).As a customer complaint (b)(4).The defective unit was not returned to the manufacturer, thus the device was not evaluated.Seven (7) units from the same manufacturing lot were retrieved from retained samples and successfully passed flow rate stability test.In addition, the batch record of the affected lot number was reviewed and no incidence similar to the one reported was detected in the controls done during the manufacturing process nor in the controls done before the product release.Therefore, the defect reported by the customer could not be reproduced.
|
| |
|
Event or Problem Description
|
|
As reported by user facility: an infusion that should have taken one hour, took five.
|
| |
|
Search Alerts/Recalls
|