• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S-ROM*STM STD,36NK,14X09X130; S-ROM HIP SYSTEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S-ROM*STM STD,36NK,14X09X130; S-ROM HIP SYSTEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 523191
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Pain (1994); Tissue Damage (2104); Discomfort (2330); Test Result (2695); No Code Available (3191)
Date of Event 05/10/2011
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Reporter is an attorney.(b)(4).
 
Event or Problem Description
Litigation alleges that the patient suffers from severe pain, discomfort in her hip, metallosis, continued medical care including revision, and difficulty walking.Update ad 12 jun 2018: wpc (b)(4) has been re-opened under (b)(4) due to the receipt of ppf and sticker sheets record.In addition to what were previously alleged, ppf alleges loosening of cup, metal wear, metallosis and elevated metal ions.Added cup and screws due to alleged loosening and stem due to alleged elevated metal ions.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.The reported event is considered one of the possible complications of joint replacement.Investigational inputs were requested as indicated per internal procedures for this failure mode.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.Without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.The device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No information received with this individual complaint indicated that a broader investigation or corrective action was necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device history lot: null.Device history batch: null.Device history review: null.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key8150487
Report Number1818910-2018-77736
Device Sequence Number1244565
Product Code JDI
UDI-Device Identifier10603295170853
UDI-Public10603295170853
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K851422
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2011
Device Catalogue Number523191
Device Lot Number2083867
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/16/2018
Supplement Date Received by Manufacturer10/04/2019
Initial Report FDA Received Date12/11/2018
Supplement Report FDA Received Date10/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
-
-