| Model Number |
PB1016 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Mitral Regurgitation (1964); Pulmonary Valve Stenosis (2024)
|
| Date of Event |
06/09/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product analysis: the device was discarded, therefore no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that ten months and twenty-four days after the implant of this transcatheter bioprosthetic pulmonary valve, into a previously placed conduit, the valve and conduit were explanted and replaced surgically with a medtronic valve.Prior to the explant, cardiac catheterization revealed a right ventricular outflow tract (rvot) pressure gradient of 30 mm hg, rvot stenosis and moderate mitral regurgitation (mr).A decrease in activity level was also reported.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided which indicated that per the physician, the conduit revision was performed due to subvalvar obstruction related to fibrous infundibulum ridge formation.Per the physician, the melody valve was not stenotic and was functioning well.If information is provided in the future, a supplemental report will be issued.
|
| |
|