| Catalog Number |
Z508G |
| Medical Device Problem Codes |
Material Separation (1562); Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Investigation summary: it was reported performance pull off suture needle pre-op.An empty opened foil, a labeled winding former with an un-dispensed suture and a needle inserted in a pink sponge were returned for analysis.During the visual inspection of the sample, the swage and attachment area were as expected; however, the suture have begun with the process of disintegration; this is the cause of the needle was detached from the strand.Also, the foil was examined for visual inspection and present holes in cavity on the back- foil package due to excessive manipulation and this condition contributed to degradation of the suture.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Per the sample condition the assignable cause of performance pull off suture needle pre-op is a suture degradation; due to the improper handling of package.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown plastic surgery on (b)(6) 2018, and suture was used.Prior to use on the patient, the suture was easily detached from the needle when taking out a needle from the tray by holding it with a needle-holder.There were no adverse consequences to the patient.Upon device evaluation, the foil was examined for visual inspection and present holes in cavity on the back- foil package due to excessive manipulation and this condition contributed to degradation of the suture.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|