| Medical Device Problem Code |
Image Display Error/Artifact (1304)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a display (dim/fading/color spectrum) issue.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may impact the user's ability to read some or all the information on the screen which may result in over or under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 04-jan-2019 with the following findings: on investigation, the display screen demonstrated a dim/fading/color spectrum issue as alleged in the initial complaint.Investigation duplicated the complaint.
|
| |
|
Additional Manufacturer Narrative
|
|
Unrelated to the original complaint, the battery compartment was noted to be cracked during the investigation.
|
| |
|
Search Alerts/Recalls
|